Test cables at
26.5 GHz and above.

Stability, loss and the real test setup.

At microwave frequencies, a test cable is part of the measurement system. Choosing one means balancing electrical stability, loss, connector quality and the way the assembly will actually move on the bench.

A cable can meet the required frequency rating and still introduce uncertainty when its shape or temperature changes after calibration. For engineers measuring small changes in a filter, amplifier or antenna, that distinction matters: the test setup must remain stable enough to separate a change in the device from a change in the connection.

This article compares five selected test-cable constructions spanning 26.5 to 70 GHz and explains how to read the specifications behind them. The comparison uses published manufacturer data; it is an editorial specification review, with no independent laboratory testing. Product selection is illustrative rather than an exhaustive market survey.

Start with the complete connection

Specify the upper test frequency, both connector interfaces and sexes, and the required reference-plane-to-reference-plane length. Include any adapters and connector savers in the configuration. The usable band depends on the complete assembly, not simply the connector name or the bulk cable’s headline capability.

Precision 3.5 mm interfaces are common in 26.5 GHz setups; higher bands often lead to 2.92 mm, 2.4 mm and 1.85 mm interfaces. Actual ratings depend on the selected parts and assembly. Mechanically compatible interfaces can also have different performance limits. Check the exact configuration rather than treating an interface family as a frequency guarantee. [1]

A picosecond becomes a visible phase change

For a simple transmission path, a change in electrical delay produces a phase change given by Δφ = 360° × f × Δt, with frequency in hertz and delay in seconds. A one-picosecond change therefore produces:

9.54°at 26.5 GHz
14.4°at 40 GHz
24.12°at 67 GHz
25.2°at 70 GHz
Calculated one-way transmission phase change for Δt = 1 ps. These are mathematical examples, not measured results or specifications for the products below. A simple reflection path traverses the cable twice.

That sensitivity explains why cable movement belongs in the measurement budget. Calibration characterizes a particular system state. Movement, connector changes or temperature drift can change that state. Calibration corrects repeatable systematic errors; it does not eliminate subsequent drift or random noise. [2]

PRODUCTS SIDE BY SIDE

Five constructions, with the conditions attached

The selected products serve different bands and handling needs. Read each figure with its qualifier; these numbers do not support a direct performance ranking.

Representative Times Clarity armored test cable and strain relief
Times Microwave Systems

Clarity 26 Armored

CLS26 26.5 GHz

Series image: Times Microwave Systems. Configured length and ends vary.

Manufacturer product page →
Junkosha MWX061 armored precision microwave test cable assembly
Junkosha

MWX0 Series

MWX061 armored 67 GHz

Junkosha catalogue image, supplied via PSI Solutions; representative assembly.

Manufacturer product page →
Representative purple GORE PHASEFLEX microwave RF test assembly
GORE

PHASEFLEX

Type 0F 70 GHz

Series image: W. L. Gore & Associates. Photo does not identify the selected Type 0F configuration.

Manufacturer product page →
Representative HUBER+SUHNER SUCOFLEX 570S test cable assemblies
HUBER+SUHNER

SUCOFLEX 500

570S · 85149825 70 GHz

Series image: HUBER+SUHNER. The pictured assembly differs from the selected 36-inch stock item.

Manufacturer product page →
Focusimple FABTest family with unarmored and armored cable assemblies
Focusimple Electronics

FABTest

FTSA armored 67 GHz

Series image: Focusimple. Multiple constructions shown; the exact 1 m FTSA assembly is not identified.

Manufacturer product page →
Manufacturer profile →
Published specifications for the selected constructions. Typical values, maxima and missing conditions are retained separately.
Specification / basisTimes Microwave SystemsCLS26JunkoshaMWX061 armoredGOREType 0FHUBER+SUHNER570S · 85149825FocusimpleFABTest FTSA armored
Selected configurationCLS26 armored construction; order exact length and terminations.MWX061 armored construction; order exact length and terminations.0F0CA0CB0360 example: 36 in / 0.91 m; 1.85 mm female–male.85149825: 914 mm / 36 in; 1.85 mm male–male.FTSA-18M18M-01.00M (T-CA21-10128): 1.0 m; 1.85 mm male–male.
Maximum frequency26.5 GHz67 GHz70 GHz70 GHz67 GHz
Interface options in cited literature3.5 mm and selected SMA options; confirm configured assembly rating.1.85 mm straight male / female.1.85 mm straight male / female.1.85 mm straight male / female; ruggedized PORT female also offered.1.85 mm straight male / female for 67 GHz configurations.
Published phase stability with flexure±1.5° typical at 26.5 GHz.±8° typical at R = 30 mm; frequency, angle and length not stated in selector.±8° typical; ±15.6° guaranteed maximum for Type 0F under the stated test.±8° typical; equivalent test geometry and frequency not stated in cited brochure.±9° typical over DC–67 GHz; geometry and assembly length not stated.
Published amplitude stability with flexure±0.035 dB typical at 26.5 GHz.Numeric value not stated in current selector / product table.±0.05 dB typical; ±0.10 dB guaranteed maximum.±0.05 dB typical.±0.10 dB typical over DC–67 GHz.
Flex-test basisShort-circuited assembly; 90° bend around 25.4 mm-radius mandrel.Current typical figure: R = 30 mm; further conditions not stated.Short-circuited assembly; 360° wrap around 57 mm-radius mandrel; test both sides with renormalization.Geometry, assembly length and test temperature not stated for these figures.Mechanical phase / amplitude figures; mandrel, angle, length and test temperature not stated.
Published loss basisNo numeric loss comparison: cited nominal per-foot and per-metre entries require reconciliation.7.3 dB/m typical at 67 GHz; per-length value, not a guaranteed finished-assembly limit.5.99 dB typical at 70 GHz for 0.91 m with straight connectors.6.48 dB maximum for the selected 914 mm, 70 GHz-rated stock assembly; product page does not separately identify the test frequency.6.84 dB listed for the selected 1 m, 67 GHz assembly; typical / maximum qualifier and temperature not stated.
Published bend radius38 mm minimum; 76 mm recommended for maximum flex life.30 mm minimum, inner-side radius in catalogue.25.4 mm minimum; static / dynamic distinction not stated.25.4 mm static minimum; separate dynamic minimum not stated.23 mm installation minimum; 47 mm repeated-bending minimum for armored FTSA.

Reading the table. Manufacturer specifications are not normalized to a common frequency, length, temperature or motion. No winner or value ranking is assigned. Service bend limits and characterization-test mandrels are different specifications.

Selection notes. Clarity 26 is used here; a manufacturer discontinuation notice covers Clarity 40/50/70/110. Its small characterization mandrel should not be treated as a permitted service radius. For MWX061, the older catalogue’s ±15° shipping criterion uses a different test from the current ±8° typical selector entry; the two are not combined. Gore’s broader family frequency claims are not attributed to Type 0F. The selected loss values have no separately stated measurement-temperature basis. [3] [4] [5] [6]

Focusimple’s FABTest addition. The selected 67 GHz FTSA assembly uses 1.85 mm interfaces and distinguishes installation from repeated-bending minima. Its typical mechanical stability figures have no published flex geometry, and the listed assembly loss has no typical / maximum qualifier. Keep those limits visible when requesting an equivalent-condition quotation. [10]

Read the flex test behind the stability number

Phase stability with flexure describes a change caused by movement; phase matching describes how closely separate assemblies’ electrical lengths agree. A closely matched pair still needs adequate stability in use. Amplitude stability similarly concerns a change in transmission or measured response during movement, rather than the assembly’s absolute insertion loss.

The selected Times and Gore specifications illustrate why the test geometry matters. Times describes a short-circuited assembly bent 90° around a 25.4 mm-radius mandrel. Gore describes a short-circuited, normalized assembly wrapped 360° around a 57 mm-radius mandrel and tested on both sides. Those are different mechanical excursions. The HUBER+SUHNER brochure does not specify an equivalent geometry for the cited stability figures. [3] [5] [6]

A short-circuit reflection method also involves a round-trip electrical path. Ask whether a reported phase excursion is a measured reflection change or an equivalent one-way transmission value, and whether “±” represents a limit about a reference or half a peak-to-peak excursion. Without those definitions, a smaller number may describe a different test.

Typical values and guaranteed limits answer different questions. Typical behavior can help identify candidates. An acceptance requirement needs an agreed limit, test method and frequency band. Missing information in a public datasheet is a question for the supplier, not evidence of inferior performance.

Compare loss at the same frequency and length

Start with the finished assembly’s insertion loss across the actual sweep. A cable-only attenuation value in dB/m does not automatically include connector transitions, and a value at 18 GHz cannot establish performance at 67 GHz. Keep typical values separate from maxima, and record the temperature basis when one is published.

For example, two matched test leads that each introduce 1.5 dB loss add 3 dB of uncorrected through-path attenuation, ignoring mismatch. Calibration can correct the stable response mathematically, but it cannot restore signal power lost before the receiver or remove the receiver’s random noise. Loss therefore remains relevant to usable measurement margin. [2]

The Gore and HUBER+SUHNER entries deliberately show the trap: both are approximately 36-inch assemblies, but the published loss figures use typical and maximum bases. Subtracting those figures would create a numerical difference without establishing a measured advantage.

Mechanical limits belong in the electrical decision

Separate a minimum routing radius from a qualified repeated-flex radius. The Times literature distinguishes a minimum from a larger radius recommended for flex life. HUBER+SUHNER explicitly labels the cited 570S radius static. A static or unlabeled minimum should not become a repeated-motion specification by assumption.

Armor can help protect a cable in a busy test area, but it also changes diameter and handling. Plan strain relief so the cable’s weight and movement are not concentrated at the connector. For a fixture that cycles continuously, qualify the complete motion path, including travel near the connector boots, rather than only a bend at the cable midpoint.

Connector care is equally practical. Inspect interfaces before mating, use the torque specified for the actual connector, and rotate the coupling nut rather than twisting the cable body. Check critical interfaces with the appropriate gauge and protect disconnected ends. Worn or contaminated connectors can undermine an otherwise careful cable selection. [7]

Qualify the cable in the motion it will see

A useful acceptance exercise is a controlled repeatability study using the intended instrument, sweep and routing. The following is an editorial qualification framework, not a standardized certification procedure:

  1. Define the configuration. Record the assembly part number, length reference, connector types, adapters, sweep, source power, IF bandwidth and temperature.
  2. Establish the reference planes. Stabilize the instrument and setup, then calibrate with the cables and connections in place using a method appropriate to the measurement. [8]
  3. Use a stable check artifact. Record a baseline trace, then move the cable through a documented, non-damaging trajectory that represents normal handling while preserving the electrical connections.
  4. Measure change across the band. Track amplitude and phase deviations relative to the baseline. Keep stationary repeat sweeps as a control so instrument drift and noise are visible.
  5. Separate mechanisms. Evaluate disconnect/reconnect repeatability in a separate sequence. Recheck after controlled temperature changes and return to the baseline condition.
  6. Set acceptance limits first. Derive them from the measurement uncertainty and decision margin, then document the results, handling conditions and a verification or replacement trigger.

A delay-only port extension is not a substitute for a full calibration at the desired planes. Instruments may offer additional loss or mismatch adjustments; use the instrument’s documented compensation model and verify its limitations. A cable’s changing response cannot be fixed permanently by entering one delay value. [9]

Choose for the workflow

A mostly stationary bench setup, a cable moved for every DUT, and a continuously cycling fixture present different requirements. Match the construction to the workflow, ask for the missing test conditions, and compare quoted assemblies on a common frequency, length and acceptance basis.

The useful purchasing question is: how much can this complete connection change during our measurement, and is that change small enough for the decision we need to make? Frequency capability gets a cable onto the shortlist. A documented, repeatable setup makes its performance meaningful.

Sources & image credits

Source review: October 28, 2026. Undated product pages are cited as specifications, not product-launch announcements. The PDF upload path does not establish its revision date. Current Junkosha values were available through indexed official content; the manufacturer-authored catalogue supplies the photographed assembly.

  1. Fluke — Choosing and using cables for RF calibration
    Technical guidance
  2. Keysight — Understanding VNA Calibrations
    Calibration, systematic errors, noise and drift
  3. Times Microwave Systems — selected CLS26 construction
    Clarity 26 Armored product page · Test & Measurement brochure, Clarity mechanical and electrical tables · Manufacturer PCN 2024-001 Rev B (DigiKey-hosted copy)
  4. Junkosha — MWX061
    Current product table · Current cable selector · Manufacturer MWX0 catalogue, MWX061 page; PSI-hosted copy and image source
  5. W. L. Gore & Associates — PHASEFLEX Type 0F
    Official datasheet, Tables 1, 3 and 5 · Family page and example assembly
  6. HUBER+SUHNER — SUCOFLEX 570S
    Stock assembly 85149825 · SUCOFLEX 500 brochure, 570 / 570S and stock-assembly tables
  7. Keysight — Connector Care
    Inspection, gauging and mating practices
  8. Rohde & Schwarz — VNA calibration methods and standards
    Reference planes and calibration
  9. Keysight — Port Extensions
    Delay and loss compensation features
  10. Focusimple Electronics — FABTest FTSA
    Official product specifications and family image · Manufacturer FABTest brochure (2021), FTSA specifications and stock-assembly table · Focusimple manufacturer profile